How to Perform a Control Solution Test

To ensure a smooth control solution testing process, please follow the instructions below to perform the test:

  1. Make sure you are using the corresponding test strips and control solution.
  2. For Ketone / Cholesterol / Lactate testing:
    • Insert the corresponding code card into the meter and verify that the number displayed on the screen matches the number on the code card, then remove it.
  3. Insert the test strip into the meter.
  4. Drip one drop of the control solution onto the cap of the bottle and use the test strip to absorb the solution
  5. Wait for the result
  6. Compare the result to the range printed on the test strip vial.

If the result falls within the specified range, this indicates that the meter and test strips are precise and accurate, allowing you to use them with confidence.

Jan 9, 2025